Issue
Date: August 23, 1999
Revision Date: March 10, 2000
Pneumercator
Company, Inc.
TMS
2000, TMS 3000
(Magnetostrictive Probe)
AUTOMATIC
TANK GAUGING METHOD
Certification
|
Leak
rate of 0.2 gph with PD = 95.0% and PFA = 5% for 2 hour
test.
Leak rate of 0.2 gph with PD = 99.8% and PFA = 0.2% for
4 hour test
Leak rate of 0.2 gph with PD = 99.9% and PFA = 0.1% for
8 hour test.
Leak rate of 0.1 gph with PD = 95.3% and PFA = 4.7% for
7 hour test.
Leak rate of 0.1 gph with PD = 95.8% and PFA = 4.2% for
8 hour test.
|
Leak
Threshold
|
0.1
gph for leak rate of 0.2 gph.
0.05 gph for leak rate of 0.1 gph.
A tank system should not be declared tight if the test
result indicates a loss that equals or exceeds this threshold.
|
Applicability
|
Gasoline,
diesel, aviation fuel.
Other liquids with known coefficients of expansion and
density may be tested after consultation with the manufacturer.
|
Tank
Capacity
|
Maximum
of 20,000 gallons.
Tank must be between 20 and 95% full.
|
Waiting
Time
|
Minimum
of 8 hours between delivery and testing.
Minimum of 20 minutes between dispensing and testing.
There must be no delivery during waiting time.
|
Test
Period
|
Minimum
of 2 hours for leak rate of 0.2 gph.
Minimum of 7 hours for leak rate of 0.1 gph.
Test data are acquired and recorded by system's computer.
Leak rate is calculated from data determined to be valid
by statistical analysis.
There must be no dispensing or delivery during test.
|
Temperature
|
Average
for product is determined by probe which contains 5 thermistors.
At least one thermistor must be submerged in product during
test.
|
Water
Sensor
|
Must
be used to detect water ingress.
Minimum detectable water level in the tank is 0.488 inch.
Minimum detectable change in water level is 0.124 inch.
|
Calibration
|
Thermistors
and probe must be checked and, if necessary, calibrated
in accordance with manufacturer's instructions.
|
Comments
|
Not
evaluated using manifolded tank systems. Therefore,
this certification is only applicable when there is
a probe used in each tank and the siphon is broken during
testing.
Tests only portion of tank containing product.
As product level is lowered, leak rate in a leaking
tank decreases (due to lower head pressure).
Consistent testing at low levels could allow a leak
to remain undetected.
EPA leak detection regulations require testing of the
portion of the tank system which routinely contains
product.
California regulations
require at least one test per month after routine product
delivery or when the tank is filled to within 10% of
the highest operating level of previous month. If the
0.1 gph test is used as a tank tightness test there
must be no water present in the backfill (groundwater
must be below the bottom of the tank), the tank must
be equipped with an overfill protection device, and
the product level in the tank must be at the overfill
protection device set point. Local agency pre-approval
is required. To use the 0.1 gph test as a monthly option
see title 23 CCR Section 2643 (b)(2).
|