Issue Date:  December 21, 2005

 

PID Analyzers, LLC
(originally listed as HNU Systems, Inc.)

Model 102 Photoionizer

OUT-OF-TANK PRODUCT DETECTOR (VAPOR-PHASE)

 

Detector:

Output type:

quantitative

Sampling frequency:

intermittent

Operating principle:

photoionization detector

Test Results:

*California regulations require vadose zone vapor monitoring to be continuous. This sensor does not meet the criteria.

unleaded gasoline

synthetic gasoline

JP-4 jet fuel

Model 102

Accuracy* (%) - average reading

85  (423 ppm)

112.8 (1125.2 ppm)

8.1 (78.2 ppm)

Detection time* (sec)

<120

11

12

Fall time* (sec)

<120

38

11

Lower detection limit (ppm)

Not Reported

2.1

5.8

*For tests conducted with 1000 ppm of test gas

Specificity Results (in addition to above):
Activated:  n-butane, n-hexane, iso-butane, 3-methylpentane, toluene

 

PID Analyzers, LLC

Evaluator:  Ken Wilcox Associates

25 Walpole Park South Drive

Tel:  (816) 443-2494

Walpole, MA 02081

Dates of Evaluation:  10/29/04

Tel:  (508) 660-5001

 

 | Home | Vendors: A - F | Vendors: G - M | Vendors: N - S | Vendors: T - Z | Method Index |

Appearance on this list is not to be construed as an endorsement by any regulatory agency nor is it any guarantee of the performance of the method or equipment.
Equipment should be installed and operated in accordance with all applicable laws and regulations.
For full details, please refer to our expanded "DISCLAIMER" page.

 
 

.